7.24
LED error messages
For Rosemount 5300 Transmitters without display, a flashing Light Emitting Diode (LED) is
used for presentation of error messages.
Figure 7-29. Rosemount 5300 Transmitter with LED
In normal operation, the LED flashes once every other second. When an error occurs, the
LED flashes a sequence that corresponds to the Code number followed by a five second
pause. This sequence is continuously repeated.
The following errors can be displayed:
Table 7-12. LED Error Codes
Example
Modem error (code 6) is displayed as the following flash sequence:
Code Error
0
Ram Failure
1
FPROM
2
HREG
4
Microwave Module
5
Display
6
Modem
7
Analog Out
8
Internal Temperature
11
Hardware
12
Measurement
14
Configuration
15
Software
3
Flashing LED
Seconds
Содержание 5300 Series
Страница 2: ......
Страница 12: ...Reference Manual 00809 0100 4530 Rev DD Contents May 2016 xii Contents ...
Страница 94: ...80 Reference Manual 00809 0100 4530 Rev DD Electrical Installation May 2016 Electrical Installation ...
Страница 162: ...148 Reference Manual 00809 0100 4530 Rev DD Operation May 2016 Operation ...
Страница 214: ...200 Reference Manual 00809 0100 4530 Rev DD Service and Troubleshooting May 2016 Service and Troubleshooting ...
Страница 310: ...Product Certifications May 2016 Reference Manual 00809 0100 4530 Rev DD 296 Product Certifications ...
Страница 358: ...Level Transducer Block May 2016 Reference Manual 00809 0100 4530 Rev DD 344 Level Transducer Block ...
Страница 431: ...5 Reference Manual 00809 0100 4530 Rev DD May 2016 ...