53
Reference Manual
00809-0100-4530, Rev DD
May 2016
Adjusting the probe length
1.
Determine
L
, the desired probe length.
2.
Determine
n
, the number of middle segments needed for the desired probe length.
See
.
3.
Calculate
Y
, the length of the bottom segment. See
and
.
4.
Continue as follows:
Length of bottom segment (Y)
Action
Continue with
step
(7)
.
Do not use the bottom segment.
Continue with
step
(5)
and cut the bottom
segment.
1. Add one extra middle segment to the
calculated
n.
2. Continue with
step (7)
.
L, desired probe length:
L
n, number of middle segments:
n
Y, length of bottom segment:
Y
Y < 0.4 in. (10 mm)
Y
≥
0.4 in. (10 mm)
Y = 31.5 in. (800 mm)