102
Reference Manual
00809-0100-4530, Rev DD
May 2016
Device Type = Link Master Device
Usage = NI-FBUS
If other host systems are connected to the bus:
Device address = Visitor
Device Type = Basic Device
Usage = NI-FBUS
2.
Start Rosemount Radar Master (RRM): from the Start menu click
Programs
>
Rosemount
>
Rosemount Radar Master
or click the RRM icon in the
MS Windows workspace.
3.
If the
National Instruments Communication Manager
server is not running, click
Yes
when RRM displays a request for starting the server.
4.
If the
Search Device
window did not appear automatically choose menu option
Device
>
Search
.
5.
In the
Search Device
window choose communication protocol F
OUNDATION
Fieldbus
(if not already selected) and click the
Start Scan
button (click the
Advanced
button if you want to specify start and stop dress).
Now RRM searches for the transmitter. After a while RRM shows the transmitters
found on the bus:
6.
Select the desired transmitter and click
OK
to connect.
In the RRM status bar verify that RRM communicates with the transmitter:
RRM communicates
with the transmitter
No communication
with the transmitter
Содержание 5300 Series
Страница 2: ......
Страница 12: ...Reference Manual 00809 0100 4530 Rev DD Contents May 2016 xii Contents ...
Страница 94: ...80 Reference Manual 00809 0100 4530 Rev DD Electrical Installation May 2016 Electrical Installation ...
Страница 162: ...148 Reference Manual 00809 0100 4530 Rev DD Operation May 2016 Operation ...
Страница 214: ...200 Reference Manual 00809 0100 4530 Rev DD Service and Troubleshooting May 2016 Service and Troubleshooting ...
Страница 310: ...Product Certifications May 2016 Reference Manual 00809 0100 4530 Rev DD 296 Product Certifications ...
Страница 358: ...Level Transducer Block May 2016 Reference Manual 00809 0100 4530 Rev DD 344 Level Transducer Block ...
Страница 431: ...5 Reference Manual 00809 0100 4530 Rev DD May 2016 ...