7.23
LCD error messages
Figure 7-28. The Rosemount 5300 Display Panel Displaying an Error Message
Table 7-11. Error Messages Displayed on the 5300 Display Panel
For more information on errors, see
.
Error message
Description
RAM FAIL
An error in the gauge data memory (RAM) has been detected
during the startup tests.
Note:
This resets the gauge automatically.
FPROM FAIL
An error in the gauge program memory (FPROM) has been
detected during the startup tests.
Note:
This resets the gauge automatically.
HREG FAIL
An error in the transmitter configuration memory (EEPROM) has
been detected. The error is either a checksum error that can be
solved by loading the default database, or a hardware error.
Note:
The default values are used until the problem is solved.
OMEM FAIL
Other memory failure.
MWM FAIL
An error in the microwave module.
DPLY FAIL
An error in the LCD.
MODEM FAIL
Modem hardware failure.
AOUT FAIL
An error in the Analog Out Module.
OHW FAIL
An unspecified hardware error has been detected.
ITEMP FAIL
An error in the internal temperature measurement.
MEAS FAIL
A serious measurement error has been detected.
CONFIG FAIL
At least one configuration parameter is outside the allowed
range.
Note:
The default values are used until the problem is solved.
SW FAIL
An error has been detected in the transmitter software.
3
Error Message
Содержание 5300 Series
Страница 2: ......
Страница 12: ...Reference Manual 00809 0100 4530 Rev DD Contents May 2016 xii Contents ...
Страница 94: ...80 Reference Manual 00809 0100 4530 Rev DD Electrical Installation May 2016 Electrical Installation ...
Страница 162: ...148 Reference Manual 00809 0100 4530 Rev DD Operation May 2016 Operation ...
Страница 214: ...200 Reference Manual 00809 0100 4530 Rev DD Service and Troubleshooting May 2016 Service and Troubleshooting ...
Страница 310: ...Product Certifications May 2016 Reference Manual 00809 0100 4530 Rev DD 296 Product Certifications ...
Страница 358: ...Level Transducer Block May 2016 Reference Manual 00809 0100 4530 Rev DD 344 Level Transducer Block ...
Страница 431: ...5 Reference Manual 00809 0100 4530 Rev DD May 2016 ...