Solid applications
Pull the probe rope through a suitable
anchoring point, e.g. a welded eye and
fasten it with two clamps. It is
recommended the probe is slack in
order to prevent high tensile loads.
The sag should be at least 1.5 in./10 ft
(1 cm/m) of the probe length.
Alternative chuck for flexible single
lead probes
Loosen the screws. Pull the probe rope
through a suitable anchoring point,
(e.g. a welded eye).
Tighten the screws. The required
torque and hex key dimensions:
4 mm wire: 5 Nm, 4 mm
6 mm wire: 10 Nm, 5 mm
1cm/m