Mounting a centering disc on rigid single probes
Note
Centering discs shall not be used with PTFE covered probes.
Rigid single lead probe (8 mm)
1.
Drill one hole using the drilling fixture (included in your shipment).
A. Drilling fixture
B. Probe
2.
Mount the bushing, centering disc, and washer at the probe end.
A. Washer
B. Centering disc
C. Bushing
3.
Insert the split pin through the bushing and the probe.
A
B
Note
Do not mount the washer if the centering disc material is PTFE,
Alloy C-276, Duplex 2205, or Alloy 400.
A
B
C