Rosemount 5300 Series with HART® to Modbus® Converter
May 2016
Reference Manual
00809-0100-4530, Rev DD
385
Rosemount 5300 Series with HART® to Modbus® Converter
6.
Connect to the transmitter and perform the necessary configuration.
7.
After completing the configuration, disconnect the RS-485 Converter, connect the
Modbus communication wires and cycle power to the transmitter
8.
Verify communication between the transmitter and the RTU is established (can
take up to 60 seconds from startup).
Note
Take the following into consideration if there are multiple Rosemount 5300 Modbus units
on the bus:
By default, the transmitters have HART address 1. It will not be possible to establish
communication on HART address 1 if several transmitters have the same address. In this
case, there are alternative solutions to establish communication:
1.
Select the Scan by Tag option in the Search Device window in RRM and enter the
HART Device Tag of the transmitter. Communication can now be established with
an individual transmitter even if several devices have the same HART address.
2.
Make sure the Rosemount 5300 Series Transmitter is alone on bus. Disconnect or
turn off power from any other devices.
J.6.2
Connect to the HART terminals
To configure the Rosemount 5300 Series Transmitter, connect the communicator or PC to
the HART terminals using a HART modem, see
. Both the
configuration tool and the RS-485 bus can be connected simultaneously. Configuration
data is sent with HART commands through the HMC to the Rosemount 5300 Transmitter
electronics, as illustrated in
. Note that the power supply must be
connected during configuration, see also
“Electrical installation” on page 378
Note
Measurement data is not updated to the Modbus Master when a configuration tool is
connected.
Содержание 5300 Series
Страница 2: ......
Страница 12: ...Reference Manual 00809 0100 4530 Rev DD Contents May 2016 xii Contents ...
Страница 94: ...80 Reference Manual 00809 0100 4530 Rev DD Electrical Installation May 2016 Electrical Installation ...
Страница 162: ...148 Reference Manual 00809 0100 4530 Rev DD Operation May 2016 Operation ...
Страница 214: ...200 Reference Manual 00809 0100 4530 Rev DD Service and Troubleshooting May 2016 Service and Troubleshooting ...
Страница 310: ...Product Certifications May 2016 Reference Manual 00809 0100 4530 Rev DD 296 Product Certifications ...
Страница 358: ...Level Transducer Block May 2016 Reference Manual 00809 0100 4530 Rev DD 344 Level Transducer Block ...
Страница 431: ...5 Reference Manual 00809 0100 4530 Rev DD May 2016 ...