May 2016
Reference Manual
00809-0100-4530, Rev DD
342
E.4
Diagnostics device errors
In addition to the BLOCK_ERR and XD_ERROR parameters, more detailed information on
the measurement status can be obtained via DIAGN_DEV_ALERT.
lists the
potential errors and the possible corrective actions for the given values. The corrective
actions are in order of increasing system level compromises. The first step should always be
to reset the gauge and then if the error persists, try the steps in
. Start with the
first corrective action and then try the second.
Table E-17. Device Errors Diagnostics
Bit
Number
Value of
DIAGN_DEV_ALERT
Description
Corrective action
0
No alarm active
0
0x00000001
Reserved
1
0x00000002
FF card to gauge comm
fault
Replace gauge
2
0x00000004
Level Measurement Failure
Verify device installation and configuration
3
0x00000008
Temperature Measurement
Failure
Check ambient temperature. If ambient
temperature ok replace the device.
4
0x00000010
Volume measurement
failure
Check volume configuration
5
0x00000020
Database Error
Load default database to the device and
reconfigure the device
6
0x00000040
HW Error
Replace device
7
0x00000080
Microwave Unit Error
Replace device
8
0x00000100
Configuration Error
Load default database to the device and
reconfigure the device
9
0x00000200
SW Error
Replace device
10
0x00000400
Invalid Strap Table
Check the Strapping Table configuration
11
0x00000800
Internal Temp Warning
Check ambient temperature at installation site
12
0x00001000
Database Warning
Check the device configuration
13
0x00002000
HW Warning
Check device installation and configuration
14
0x00004000
Microwave Unit Warning
Check device installation and configuration
15
0x00008000
Configuration Warning
Check the device configuration
16
0x00010000
SW Warning
Check device installation and configuration
17
0x00020000
Simulation Mode
Use Simulation Method under advanced
configuration tool to get the device out of
Simulation Mode
18
0x00040000
Volume Range Warning
Verify Strapping Table
19
0x00080000
Software Write Protected
20
0x00100000
Vapor Compensation Not
Calibrated
21
0x00200000
Vapor DC Estimation
Limited
Содержание 5300 Series
Страница 2: ......
Страница 12: ...Reference Manual 00809 0100 4530 Rev DD Contents May 2016 xii Contents ...
Страница 94: ...80 Reference Manual 00809 0100 4530 Rev DD Electrical Installation May 2016 Electrical Installation ...
Страница 162: ...148 Reference Manual 00809 0100 4530 Rev DD Operation May 2016 Operation ...
Страница 214: ...200 Reference Manual 00809 0100 4530 Rev DD Service and Troubleshooting May 2016 Service and Troubleshooting ...
Страница 310: ...Product Certifications May 2016 Reference Manual 00809 0100 4530 Rev DD 296 Product Certifications ...
Страница 358: ...Level Transducer Block May 2016 Reference Manual 00809 0100 4530 Rev DD 344 Level Transducer Block ...
Страница 431: ...5 Reference Manual 00809 0100 4530 Rev DD May 2016 ...