Prerequisites
Before clicking the Trim Near Zone button, ensure that:
There is product in the tank.
The product level is below the Near Zone region (0-3.3 ft, 0-1 m) below the Upper
Reference Point).
The ambient temperature is within 104
°
F (40
°
C) of the expected ambient
operating temperature.
Procedure
1.
In RRM, select
Setup
>
Advanced
and click the
Near Zone
tab.
The function can also be reached from
Device specific setup
in the
Guided Setup
(if the command is recommended).
2.
Click
Trim Near Zone
.
3.
Select
Trim near zone
and click
OK
.
4.
Click
OK
in the appearing dialog.
5.
When near zone trimming is complete, restart the device.
Select
Tools
>
Restart Device
.
6.
Restart the transmitter.
To reset the Trim Near Zone function, do the following:
1.
Select
Setup
>
Advanced
and click the
Near Zone
tab.
2.
Click
Trim Near Zone
.
3.
Select
Reset near zone trimming to factory settings
and click
OK
.
Field Communicator
In a Field Communicator, the Trim Near Zone is reached with the [2, 1, 7, 2] HART sequence
(if the command is recommended) or from the [2, 7, 1] HART sequence.
Wait one minute and restart the transmitter.
Содержание 5300 Series
Страница 2: ......
Страница 12: ...Reference Manual 00809 0100 4530 Rev DD Contents May 2016 xii Contents ...
Страница 94: ...80 Reference Manual 00809 0100 4530 Rev DD Electrical Installation May 2016 Electrical Installation ...
Страница 162: ...148 Reference Manual 00809 0100 4530 Rev DD Operation May 2016 Operation ...
Страница 214: ...200 Reference Manual 00809 0100 4530 Rev DD Service and Troubleshooting May 2016 Service and Troubleshooting ...
Страница 310: ...Product Certifications May 2016 Reference Manual 00809 0100 4530 Rev DD 296 Product Certifications ...
Страница 358: ...Level Transducer Block May 2016 Reference Manual 00809 0100 4530 Rev DD 344 Level Transducer Block ...
Страница 431: ...5 Reference Manual 00809 0100 4530 Rev DD May 2016 ...