GX-6000 Operator’s Manual
Chapter 4: Calibration Mode • 85
8 . Press and release the POWER/ENTER button. The target gas selection screen will
display. This screen allows you to select the gas you want to use for calibration.
Isobutylene will always be displayed on the top line. The second line will show the
gas that is currently selected in the PID Gas Name Screen in Display Mode. In the
example below, acetone is the currently selected gas. If isobutylene is the selected
gas, then isobutylene will appear twice in this list.
9 . Use the
▲
AIR and SHIFT
▼
(PANIC) buttons to move the cursor next to the gas you
want to use for calibration. Continue to Step 10.
10 . Press and release the POWER/ENTER button to proceed to the Single Calibration
Apply Gas Screen.
APPLYGAS
and the gas reading will be flashing.
The screen shown below is for the CH
4
channel. If you are calibrating a different
channel, your screen will look slightly different.
11 . Connect the tubing from the demand flow regulator to the rigid tube on the probe.
Allow the GX-6000 to draw gas for one minute.
12 . If necessary, adjust the reading to match the cylinder concentration with the
▲
AIR
and SHIFT
▼
(PANIC) buttons. If the reading cannot be adjusted to match the
cylinder concentration, see “Troubleshooting” on page 100.
SINGLE
GAS SELECT
Isobutylene
Acetone
ESCAPE
APPLYGAS
CH4
%LEL
0