GX-6000 Operator’s Manual
Appendix D: Using the GX-6000 in Bar Hole Mode • 179
3 . Use the
▲
AIR and SHIFT
▼
(PANIC) buttons to scroll to the desired station ID and
press and release POWER/ENTER. The unit will return to the Bar Hole Measuring
Screen.
4 . Take the GX-6000 to the bar hole that will be tested.
5 . Insert the probe into the bar hole and press and release the POWER/ENTER button.
The pump will turn on and the sample period will begin with the sample period
counting down in seconds in the lower middle portion of the display. “MEASURE”
will flash below the countdown. The combustible gas channel will automatically
switch between %LEL or %vol based on the level of gas present. The combustible
gas channel’s reading will also be displayed in graph format on the bottom of the
screen. If you wish to cancel the measurement, press and release the RESET button.
4:07
CH4
O2
0
%LEL
20.9
%
VALVE 3
4:07
CH4
O2
50
%LEL
20.9
%
30 SEC
MEASURE