136 • Appendix A: Maintenance Mode
GX-6000 Operator’s Manual
9 . If Maintenance Mode’s
PASSWORD
is set to
ON
(factory setting), enter the
password by using the
▲
AIR and SHIFT
▼
(PANIC) buttons to select each password
number and then pressing and releasing the POWER/ENTER button to enter it and
move on to the next number until all of the numbers are entered. The password is
0006.
10 . If you enter an incorrect password, an error screen will appear.
You must turn the unit off and reenter Maintenance Mode using the correct
password. Continue to Step 12.
11 . If Maintenance Mode’s
PASSWORD
is set to
OFF
, press and release the POWER/
ENTER button to enter Maintenance Mode. There is no need to enter a password.
The password screen appears even if
PASSWORD
is set to
OFF
so that a special
factory mode can be accessed, if necessary.
12 . The main menu displays. It displays eight menu items at a time.
13 . Use the
▲
AIR or SHIFT
▼
(PANIC) button to move the cursor up and down through
the menu items. Additional items are available above and below the items currently
displayed on the screen. They will appear as you scroll farther up or down the current
list using the
▲
AIR or SHIFT
▼
(PANIC) buttons.
PASSWORD
PASSWORD
ERROR
MAINTE
MAINTENANCE
DATE
DATE FORMAT
GAS CALIBRATION
CAL SETTING
BUMP SETTING
ALARM SETTING
PUMP OFF DISP
ID DISPLAY