242 • Appendix J: Creating a PID User List
GX-6000 Operator’s Manual
6 . To reorder the gases in the PID User List:
a. Select a gas in the PID User List.
b. Click
Up
and
Down
to move the selected gas up or down. The gas order shown
in the PID User List is how the gases will appear when you access the User List
in the GX-6000’s PID Gas List Screen.
7 . If you want to save the PID User List, making it easy to put the load the same PID
User List into multiple instruments, click
Save
. Each Save operation only saves the
currently displayed list. If you want to create PID User Lists for each sensor type
(10.6 eV, 10.0 eV, or 11.7 eV), each PID User List must be saved separately.
8 . Click
Upload
to save the PID User List to the GX-6000.
9 . If necessary, repeat Step 3 - Step 8 to edit the PID User Lists for the other PID sensor
types (10.6 eV, 10.0 eV, or 11.7 eV). Each sensor type has its own PID User List.
10 . Click
Detector off
to turn off the GX-6000.
11 . Click the red X to close the PID Gas List Editor program. A window will pop up
asking if you want to keep the current list.
Clicking
Yes
makes the currently displayed list the default list. This means when you
reopen the PID Gas List Editor Program, the currently displayed list will appear
instead of the default list that appeared when you opened the program for the first
time.
Clicking
No
leaves the PID Gas List Editor Program’s default lists alone. When you
reopen the program, you will see the default list and not the currently displayed list.