GX-6000 Operator’s Manual
Appendix H: 10.0 eV/Benzene PID Sensor • 227
11 . An automatically-determined, temperature-based countdown is displayed at the
bottom of the screen. The number to the left of the countdown (“5” in the example
below) is the temperature code.
12 . At the end of the countdown, use the
▲
AIR or SHIFT
▼
(PANIC) button to adjust
the reading to match the concentration listed on the calibration cylinder.
13 . Press and release the POWER/ENTER button.
14 . If the calibration passes, the screen will indicate PASS and will return you to the
TUBE CAL Menu.
If the reading cannot be adjusted to match the cylinder concentration, see
“Troubleshooting” on page 100.
15 . Use the SHIFT
▼
(PANIC) button to place the cursor next to
ESCAPE
, then press
and release the POWER/ENTER button. You will be prompted to remove the filter
tube and tube holder.
16 . Remove the tube holder from the GX-6000’s inlet fitting.
17 . Press and release the POWER/ENTER button. You will return to the Calibration
Mode Menu.
18 . Use the SHIFT
▼
(PANIC) button to place the cursor next to the
NORMAL MODE
menu option, then press and release the POWER/ENTER button to return to
Measuring Mode.
19 . Detach the calibration tubing from the tube holder’s inlet fitting.
20 . Unscrew the demand flow regulator from the calibration cylinder.
APPLYGAS
BNZ
5 45SEC
0.00
ppm
TUBE CAL
REMOVE HOLDER
YES :ENTER
NO :DISP