GX-6000 Operator’s Manual
Chapter 4: Calibration Mode • 77
15 . Press and release the POWER/ENTER button. The first of the Gas Select Screens,
the standard four sensors screen, will be displayed. In addition to the standard four
sensors screen, there is a separate screen for each installed smart sensor, and an
ESCAPE
screen that will take you back to the Auto Cal menu. Use the
▲
AIR and
SHIFT
▼
(PANIC) buttons to scroll through the different screens.
16 . With the standard four sensors screen displayed, press and release the POWER/
ENTER button to proceed to the Calibration In Process Screen with
APPLYGAS
and the gas readings flashing.
If you do not want to proceed with the calibration, press and release the DISP/LOCK
button to return to the Gas Select Screen.
If you do want to continue with the calibration, proceed to the next step.
AUTO
GAS SELECT
CH4 50 %LEL
O2 12.0 %
H2S 25.0 ppm
CO 50 ppm
AUTO
GAS SELECT
VOC 100 ppm
AUTO
GAS SELECT
ESCAPE
▲
AIR or
SHIFT
▼
(PANIC)
▲
AIR or
SHIFT
▼
(PANIC)
▲
AIR or
SHIFT
▼
(PANIC)
APPLYGAS
CH4 O2
%LEL %
H2S CO
ppm ppm
0 20.9
0.0 0