188 • Appendix F: ESS-03 Toxic Sensors
GX-6000 Operator’s Manual
3 . After you have completed Step 30 on page 81, the Auto Cal Menu will be displayed.
Continue with Step 4 below.
4 . If you were using a 5-gas (standard 4-gas + ESS-03) calibration cylinder to calibrate
the standard 4 sensors, continue to Step 5.
If you were using a 5-gas (standard 4-gas + high range PID) calibration cylinder or a
4-gas calibration cylinder/high range PID-only calibration cylinder combination to
calibrate the standard 4 sensors and/or the high range PID sensor:
a. Remove the tubing from the rigid tube on the probe.
b. Unscrew the 5-gas (standard 4-gas + high range PID), 4-gas cylinder, or high
range PID cylinder from the regulator.
c. Screw the ESS-03 calibration cylinder onto the demand flow regulator. Continue
with Step 6.
5 . If you were using a 5-gas (standard 4-gas + ESS-03) calibration cylinder to calibrate
the standard 4 sensors:
a. Leave the tubing connected to the rigid tube on the probe. Continue with Step 6.
6 . Use the
▲
AIR and SHIFT
▼
(PANIC) buttons to move the cursor next to
GAS
SELECT
.
AUTO
AUTO CAL
CONCENTRATION
GAS SELECT
ESCAPE
AUTO
AUTO CAL
CONCENTRATION
GAS SELECT
ESCAPE