GX-6000 Operator’s Manual
Chapter 6: Maintenance • 115
8 . Place the sensor gasket and filters back over the sensors ensuring that the sensor
gasket seals with the sensors and the flow fittings. The flow fitting connections need
to be facing up and cannot be skewed sideways. The gasket must be pushed down in
the center and secured under the gasket aligning tabs.
Failure to install the sensor gasket appropriately may result in inaccurate gas
readings.
Figure 26: Sensor Gasket Seating
9 . Insert the flow chamber back into the instrument.
10 . Tighten the flow chamber’s four screws that were loosened in Step 2.
11 . Start up the GX-6000 by pressing and briefly holding the POWER/ENTER button.
12 . To verify that the sensor gasket was inserted properly:
a. Plug the inlet with your finger.
b. Verify that the GX-6000 goes into low flow alarm. If the GX-6000 does not go
into low flow alarm, turn the instrument off and attempt to seat the sensor gasket
again.
13 . Calibrate the new sensors as described in “Chapter 4: Calibration Mode” on page 69.
Gasket Aligning Tabs
Flow Fitting Connections