GX-6000 Operator’s Manual
Chapter 3: Operation • 27
instructions to calibrate the GX-6000. When you are done with the calibration
and exit Calibration Mode, the unit will begin the startup sequence. If the
calibration was successful, the screen above will not appear again until the unit is
due for calibration. If the calibration was not successful, the screen above will
again appear in the startup sequence.
To continue without performing a calibration, press and release the RESET
button.
•
If the unit is due for calibration and
CAL EXPRD
is set to
CANNOT USE
, then
the following screen displays and the buzzer sounds in a double pulsing pattern.
The GX-6000 cannot be used until a successful calibration has been performed.
Press and release the POWER/ENTER button to enter Calibration Mode. See
“Chapter 4: Calibration Mode” on page 69 for instructions to calibrate the GX-
6000.
NOTE:
In this situation, even if the User password function has been turned on, no
password is required to perform a calibration.
When you are done with the calibration and exit Calibration Mode, the unit will
begin the startup sequence. If the calibration was successful, the screen above
will not appear again until the unit is due for calibration. If the calibration was
not successful, the screen above will again appear in the startup sequence.
4:07
CAL DATE PAST
CANNOT USE
CAL MODE : ENTER