GX-6000 Operator’s Manual
Appendix H: 10.0 eV/Benzene PID Sensor • 225
17 . Use the SHIFT
▼
(PANIC) button to place the cursor next to
ESCAPE
, then press
and release the POWER/ENTER button. You will be prompted to remove the filter
tube and tube holder.
18 . Remove the tube holder from the GX-6000’s inlet fitting.
19 . Press and release the POWER/ENTER button. You will return to the Calibration
Mode Menu.
20 . Use the SHIFT
▼
(PANIC) button to place the cursor next to the
NORMAL MODE
menu option, then press and release the POWER/ENTER button to return to
Measuring Mode.
21 . Detach the calibration tubing from the tube holder’s inlet fitting.
22 . Unscrew the demand flow regulator from the calibration cylinder.
Single TUBE CAL
1 . To enter Calibration Mode, while in Measuring Mode, press and hold the SHIFT
▼
(PANIC) button, then press the DISP/LOCK button and release both buttons.
2 . Use the SHIFT
▼
(PANIC) button to place the cursor next to
TUBE CAL
and then
press and release the POWER/ENTER button.
TUBE CAL
REMOVE HOLDER
YES :ENTER
NO :DISP
CAL
CAL MODE
AIR CAL
AUTO CAL
SINGLE CAL
TUBE CAL
CAL CODE
NORMAL MODE