140 • Appendix A: Maintenance Mode
GX-6000 Operator’s Manual
Updating the Calibration Expired Setting (CAL EXPRD)
This item defines what indication is given during start up when calibration is due and
CAL REMINDER
is set to
ON
.
With
CAL EXPRD
set to
CONFIRM TO USE
(factory setting), the GX-6000 will give
an indication at start up if calibration is past due and will require the user to decide
whether to perform a calibration or continue and use the GX-6000 without calibrating.
Press and release DISP/LOCK to continue without calibrating or POWER/ENTER to
perform a calibration.
With
CAL EXPRD
set to
CANNOT USE
, if the unit is due for calibration, the GX-6000
will give an indication at start up that calibration is past due and will prompt you to press
and release POWER/ENTER to enter Calibration Mode and perform a calibration. Using
any other button will have no effect. A successful calibration must be performed in order
to use the instrument.
NOTE:
Even if the password function is turned on in
PASSWORD
, no password will
be required to perform a calibration during startup.
With
CAL EXPRD
set to
NO EFFECT,
the GX-6000 will give an indication at startup
that calibration is past due. If desired, the POWER/ENTER button can be pressed to enter
Calibration Mode and perform a calibration but it is not necessary to acknowledge the
calibration due indication. The warm-up sequence will continue on its own.
1 . From the
CAL SETTING
screen, place the cursor next to
CAL EXPRD
.
2 . Press and release POWER/ENTER. The Calibration Expired Screen appears.
3 . Use
▲
AIR or SHIFT
▼
(PANIC) to display the desired setting.
4 . Press and release POWER/ENTER to save the setting and return to the
CAL
SETTING
menu.
Updating the Calibration Check Gas Setting (CAL CHECK GAS)
This item defines which sensors are used to update the calibration reminder screen. With
CAL CHECK GAS
set to
4 GAS
(factory setting), the calibration dates for only the
standard 4 sensors (LEL, O2, H2S, and CO) are used to determine if a calibration is due.
With
CAL CHECK GAS
set to
ALL GAS
, the calibration dates for the standard 4
sensors along with the PID sensor and any other installed smart sensors are used to
determine if a calibration is due.
1 . From the
CAL SETTING
screen, place the cursor next to
CAL CHECK GAS
.
2 . Press and release POWER/ENTER. The Calibration Reminder Screen appears.
3 . Use
▲
AIR or SHIFT
▼
(PANIC) to display the desired setting.
4 . Press and release POWER/ENTER to save the setting and return to the
CAL
SETTING
menu.
Returning to the Maintenance Mode Menu (ESCAPE)
1 . From the
CAL SETTING
screen, place the cursor next to
ESCAPE
.
2 . Press and release POWER/ENTER. The instrument will return to the main menu.