216 • Appendix H: 10.0 eV/Benzene PID Sensor
GX-6000 Operator’s Manual
d. At the end of the STEL countdown, the STEL reading displays.
e. To repeat the STEL measurement, press and release POWER/ENTER.
f. To return to Measuring Mode, press and release the DISP/LOCK button. You will
be prompted to remove the tube holder from the instrument’s inlet fitting.
g. Remove the tube holder from the GX-6000’s inlet fitting.
h. Press and release the POWER/ENTER button.
TUBE
BNZ
STEL
REPEAT ?
YES :ENTER
NO :DISP
0.00
ppm
TUBE
REMOVE HOLDER
YES :ENTER
NO :DISP