GX-6000 Operator’s Manual
Appendix H: 10.0 eV/Benzene PID Sensor • 211
Benzene Select Mode
Benzene Select Mode allows the user to monitor for benzene in the 0-50.00 ppm range.
Readings taken in Benzene Select Mode are saved to the instrument’s memory. See
“Viewing Benzene Select Mode Data” on page 217 for a description of viewing data.
NOTE:
A PID-003L 10.0 eV/benzene sensor must be installed in order for the Benzene
Select Mode screen to appear.
A tube holder and filter tube are necessary for monitoring an area in Benzene Select
Mode.
Be sure to perform an appropriate calibration for Benzene Select Mode use as described in
“Important Calibration Information” on page 219 before continuing.
WARNING:
If you take the GX-6000 to an area that is 10 or 15 degrees different
than the previous area, allow the instrument 45 minutes to acclimate
before taking any readings. Failure to do so may cause the reading to
be inaccurate.
WARNING:
There are no alarm indications in Benzene Select Mode.
1 . Enter Display Mode by pressing the DISP/LOCK button while in Measuring Mode.
The Benzene Select Mode Screen appears.
2 . Press and release the POWER/ENTER button to enter Benzene Select Mode.
DISP
BENZENE SELECT
YES :ENTER
NO :DISP
EXIT :RESET