GX-6000 Operator’s Manual
Chapter 6: Maintenance • 105
7 . Install a new battery pack. If the battery pack release latch moved back into its closed
position, slide the latch to the right with one hand and insert a new battery pack with
the other.
Figure 15: Installing the Battery Pack
8 . Release the battery pack release latch. If it does not return to its closed position
automatically, push the latch to the left and toward the instrument until it is flush
with the bottom of the instrument.
9 . Reinstall the rubber boot, if desired.
Recharging the Lithium Ion Battery Pack in the Instrument
CAUTION:
To be used only with lithium ion battery pack p/n 49-1619. Charge only
with RKI charger model BC-6000, p/n 49-2182, or RKI charger model BC-
6000DC, p/n 49-2183. Use of other rechargeable batteries or chargers or
charging of other rechargeable batteries in the GX-6000 will void the
warranty.
PRUDENCE:
Pour être utilisé uniquement avec une batterie au lithium-ion p/n 49-1619.
Charge uniquement en fonction du modèle de chargeur RKI BC-6000, p/n
49-2182, ou modèle de chargeur BC-6000DC RKI, p/n 49-2183.
L’utilisation d’autres piles ou chargeurs rechargeables ou charger d’autres
batteries recharegables dans le GX-6000 annule la garantie.
The charger can be mounted to the wall using DIN rail, if desired. See “DIN Rail
Mounting Assembly” on page 22 for instructions.
1 . Make sure the GX-6000 is off.
2 . Confirm that the power adapter’s DC output cord is plugged into the charger.
2.
1. Be Sure Battery Pack Release
Latch is Pushed to the Right
Push New Battery Pack Down