GX-6000 Operator’s Manual
Appendix A: Maintenance Mode • 143
2 . Press and release POWER/ENTER. The current Cal Time value will begin to flash.
3 . Use
▲
AIR or SHIFT
▼
(PANIC) to display the desired setting.
4 . Press and release POWER/ENTER to save the setting. The value will stop flashing.
Updating the AUTO CAL Setting
With
AUTO CAL
set to
ON
(factory setting), if a bump test fails, the unit will
automatically begin a calibration. If
AUTO CAL
is set to
OFF
and a bump test fails,
nothing will happen.
1 . From the
BUMP PARAMETERS
screen, place the cursor next to
AUTO CAL
.
2 . Press and release POWER/ENTER. The current Auto Cal setting will begin to flash.
3 . Use
▲
AIR or SHIFT
▼
(PANIC) to display the desired setting.
4 . Press and release POWER/ENTER to save the setting. The setting will stop flashing.
Returning to the BUMP SETTINGS Menu (ESCAPE)
1 . From the
BUMP PARAMETERS
screen, place the cursor next to
ESCAPE
.
2 . Press and release POWER/ENTER. The instrument will return to the
BUMP
SETTING
screen.
Updating the Bump Reminder Setting (BUMP REMINDER)
With
BUMP REMINDER
set to
ON
, the GX-6000 will give an indication at start up if it
is due for bump testing. The type of indication will depend on the
BUMP EXPRD
setting
(see “Updating the Bump Expired Setting (BUMP EXPRD)” on page 144).
NOTE:
If
BUMP REMINDER
is set to
ON
but
BUMP DISP
is set to
OFF
, the
instrument will alert you if a bump test is due but the bump test item will not
appear in Calibration Mode.
With
BUMP REMINDER
set to
OFF
(factory setting), the GX-6000 will not give an
indication at start up if it is due for bump testing.
1 . From the
BUMP SETTING
screen, place the cursor next to
BUMP REMINDER
.
2 . Press and release POWER/ENTER. The current Bump Reminder setting will be
displayed.
3 . Use
▲
AIR or SHIFT
▼
(PANIC) to display the desired setting.
4 . Press and release POWER/ENTER to save the setting. The instrument will return to
the
BUMP SETTING
screen.
Updating the Bump Interval (BUMP INTERVAL)
This setting defines the amount of time between bump tests. The time can be set in 1 day
increments. The minimum setting is 0 days and the maximum setting is 30 days. The
factory setting is 30 days.
1 . From the
BUMP SETTING
screen, place the cursor next to
BUMP INTERVAL
.
2 . Press and release POWER/ENTER. The current Bump Interval setting will be
displayed.
3 . Use
▲
AIR or SHIFT
▼
(PANIC) to display the desired setting.
4 . Press and release POWER/ENTER to save the setting. The instrument will return to
the
BUMP SETTING
screen.