GX-6000 Operator’s Manual
Appendix A: Maintenance Mode • 141
Updating Bump Test Settings (BUMP SETTING)
The
BUMP SETTING
menu item in Maintenance Mode has a sub menu with 7 menu
items:
BUMP DISP
,
BUMP PARAMETERS
,
BUMP REMINDER
,
BUMP
INTERVAL
,
BUMP EXPRD
,
BUMP CHECK GAS
, and
ESCAPE
.
1 . From the main menu, place the cursor next to
BUMP SETTING
.
2 . Press and release POWER/ENTER. The
BUMP SETTING
menu appears.
Updating the Bump Display Setting (BUMP DISP)
With
BUMP DISP
set to
ON
, the
BUMP TEST
menu item will appear in Calibration
Mode.
NOTE: BUMP DISP
must be set to
ON
to have the GX-6000’s bump test date reset
after a successful bump test in an SDM-6000.
With
BUMP DISP
set to
OFF
(factory setting), the
BUMP TEST
menu item will not
appear in Calibration Mode.
1 . From the
BUMP SETTING
screen, place the cursor next to
BUMP DISP
.
2 . Press and release POWER/ENTER. The Bump Display Screen appears.
3 . Use
▲
AIR or SHIFT
▼
(PANIC) to display the desired setting.
4 . Press and release POWER/ENTER to save the setting and return to the
BUMP
SETTING
menu.
MAINTE
BUMP SETTING
BUMP DISP
BUMP PARAMETERS
BUMP REMINDER
BUMP INTERVAL
BUMP EXPRD
BUMP CHECK GAS
ESCAPE