Appendix A: Maintenance Mode • 155
Updating Datalogging Parameters (LOG SETTING)
The
LOG SETTING
menu item in Maintenance Mode has a sub menu with 4 menu
items:
LOG INTERVAL
,
LOG CLEAR DISP
,
LOG OVER WRITE
, and
ESCAPE
.
1 . From the main menu, place the cursor next to
LOG SETTING
.
2 . Press and release POWER/ENTER. The
LOG SETTING
menu appears.
Updating the Data Log Interval Setting (LOG INTERVAL)
This setting indicates how often the GX-6000 saves readings to the data logger. The
following interval times can be selected: 10 minutes, 5 minutes (factory setting), 3
minutes, 2 minutes, 1 minute, 30 seconds, 20 seconds, or 10 seconds.
1 . From the
LOG SETTING
screen, place the cursor next to
LOG INTERVAL
.
2 . Press and release POWER/ENTER. The Log Interval Screen appears.
3 . Use
▲
AIR or SHIFT
▼
(PANIC) to display the desired setting.
4 . Press and release POWER/ENTER to save the setting and return to the
LOG
SETTING
screen.
Updating the Log Clear Display Setting (LOG CLEAR DISP)
With
LOG CLEAR DISP
set to
ON
(factory setting), the LOG CLEAR Screen appears
in Display Mode (see “LOG CLEAR Screen” on page 55).
With
LOG CLEAR DISP
set to
OFF
, the LOG CLEAR Screen does not appear in
Display Mode.
1 . From the
LOG SETTING
screen, place the cursor next to
LOG CLEAR DISP
.
2 . Press and release POWER/ENTER. The Log Clear Display Screen appears.
3 . Use
▲
AIR or SHIFT
▼
(PANIC) to display the desired setting.
4 . Press and release POWER/ENTER to save the setting and return to the
LOG
SETTING
screen.
Updating the Data Log Overwrite Setting (LOG OVER WRITE)
With
LOG OVER WRITE
set to
ON
(factory setting), the GX-6000 writes over the
oldest data with new data when the data logger memory is full.
With
LOG OVER WRITE
set to
OFF
, the GX-6000 stops saving data to the data logger
when the data logger memory is full.
1 . From the
LOG SETTING
screen, place the cursor next to
LOG OVER WRITE
.
MAINTE
LOG SETTING
LOG INTERVAL
LOG CLEAR DISP
LOG OVER WRITE
ESCAPE