28 • Chapter 3: Operation
GX-6000 Operator’s Manual
•
If the unit is due for calibration and
CAL EXPRD
is set to
NO EFFECT
, then
the following alert screen displays and the buzzer sounds in a double pulsing
pattern.
If you want to enter Calibration Mode, press and release the POWER/ENTER
button. If you do not want to enter Calibration Mode, wait for the instrument to
continue with its startup sequence.
5 . If
BUMP REMINDER
is set to
ON
(factory setting is
OFF
), the screen that appears
next depends on how
BUMP EXPRD
is set in Maintenance Mode (see “Updating
the Bump Expired Setting (BUMP EXPRD)” on page 144).
•
If the unit is due for bump testing and
BUMP EXPRD
is set to
CONFIRM TO
USE
, then the following screen displays and the buzzer sounds in a double
pulsing pattern.
To perform a bump test, press and release the POWER/ENTER button. The GX-
6000 will enter Calibration Mode and the LCD will show the Calibration Mode
main menu. See “Chapter 4: Calibration Mode” on page 69 for instructions to
bump test the GX-6000. When you are done with the bump test and exit
Calibration Mode, the unit will begin the startup sequence. If the bump test was
successful, the screen above will not appear again until the unit is due for bump
testing. If the bump test was not successful, the screen above will again appear in
the startup sequence.
To continue without performing a bump test, press and release the RESET
button.
4:07
CAL DATE PAST
NO EFFECT
CAL MODE : ENTER
4:07
BUMP DATE PAST
CONFIRM TO USE
CAL MODE : ENTER
NO : RESET