240 • Appendix J: Creating a PID User List
GX-6000 Operator’s Manual
Appendix J: Creating a PID User List
Overview
This appendix explains how to set up a User List for the GX-6000’s PID sensor(s). This
User List appears in the PID Gas Name Screen in the GX-6000’s Display Mode. The PID
User List is not related to the GX-6000’s User ID list.
Materials
In order to set up the GX-6000’s PID User List, you will need:
•
IrDA/USB cable
•
GX-6000
•
PID Gas List Editor program
•
computer with Windows 7, Windows 8, or Windows 10
Procedure
Installing the PID Gas List Editor Program
1 . Double click the setup.exe file.
2 . Follow the Setup Wizard prompts to install the program.
Connecting a GX-6000
1 . Plug your IrDA/USB cable into your computer.
2 . Launch the PID Gas List Editor Program (appears as “GX-6000PID” in the program
list).
3 . Align the IrDA/USB cable with the GX-6000’s IrDA window (located directly below
the POWER/ENTER button).
4 . Press and release the POWER/ENTER button to turn on the GX-6000.
5 . When a connection is established, the GX-6000’s LCD will say “PC TRANSMIT”.
The PID Gas List Editor Program does not show any connection indication.