GX-6000 Operator’s Manual
Appendix A: Maintenance Mode • 139
Updating Calibration Settings (CAL SETTING)
The
CAL SETTING
menu item in Maintenance Mode has a sub menu with 5 menu
items:
CAL REMINDER
,
CAL INTERVAL
,
CAL EXPRD
,
CAL CHECK GAS
, and
ESCAPE
.
1 . From the main menu, place the cursor next to
CAL SETTING
.
2 . Press and release POWER/ENTER. The
CAL SETTING
menu appears.
Updating the Calibration Reminder Setting (CAL REMINDER)
With
CAL REMINDER
set to
ON
(factory setting), the GX-6000 will give an indication
at start up if it is due for calibration. The type of indication will depend on the
CAL
EXPRD
setting (see “Updating the Calibration Expired Setting (CAL EXPRD)” on
page 140).
With
CAL REMINDER
set to
OFF
, the GX-6000 will not give an indication at start up if
it is due for calibration.
1 . From the
CAL SETTING
screen, place the cursor next to
CAL REMINDER
.
2 . Press and release POWER/ENTER. The Calibration Reminder Screen appears.
3 . Use
▲
AIR or SHIFT
▼
(PANIC) to display the desired setting.
4 . Press and release POWER/ENTER to save the setting and return to the
CAL
SETTING
menu.
Updating the Calibration Interval (CAL INTERVAL)
This setting defines the amount of time between calibrations. The time can be set in 1 day
increments. The minimum setting is 1 day and the maximum setting is 365 days. The
factory setting is 90 days.
1 . From the
CAL SETTING
screen, place the cursor next to
CAL INTERVAL
.
2 . Press and release POWER/ENTER. The Set Calibration Interval Screen appears.
3 . Use
▲
AIR or SHIFT
▼
(PANIC) to display the desired setting.
Tip:
Press and hold
▲
AIR or SHIFT
▼
(PANIC) to rapidly scroll through the settings.
4 . Press and release POWER/ENTER to save the setting and return to the
CAL
SETTING
menu.
MAINTE
CAL SETTING
CAL REMINDER
CAL INTERVAL
CAL EXPRD
CAL CHECK GAS
ESCAPE