GX-6000 Operator’s Manual
Chapter 3: Operation • 45
Responding to Clock Failure Alarms
A clock failure alarm occurs if the unit’s internal clock malfunctions.
1 . Press and release the RESET button to continue into Measuring Mode.
CAUTION:
There will be no datalogging function if you operate the instrument after a
clock failure.
2 . Attempt to change the date using the
DATE
menu item in Maintenance Mode. See
“Setting the Date and Time (DATE)” on page 137.
3 . If the date cannot be set correctly, contact RKI Instruments, Inc. as soon as possible.
Responding to System Failure Alarms
1 . If a system failure occurs, the system failure screen will display an error code as
shown below:
2 . The error code meanings are shown in the table below:
Table 8: Error Code Explanation
Error Code
Explanation
000
ROM failure
010
RAM failure
021
FRAM failure
022
FLASH memory failure
4:07
FAIL CLOCK
4:07
FAIL SYSTEM
0 2 2