GX-6000 Operator’s Manual
Appendix H: 10.0 eV/Benzene PID Sensor • 231
7 . If you are replacing your old tube with a tube from a new box, compare the
calibration code listed on the old box to the calibration code listed on the new box.
a. If the calibration codes match, there is no need to do anything else.
b. If the calibration codes do not match, you must either perform a TUBE CAL
using a tube from the new box (see “Performing a TUBE CAL” on page 221) or
you must enter the new calibration code into CAL CODE (see “Entering a CAL
CODE” on page 228), depending on what method of calibration you used with
your old box of filter tubes. See “Calibrating the 10.0 eV/Benzene PID Sensor”
on page 219 for a description of calibration methods.
8 . Tubes do not contain cadmium, mercury, or chrome and may be disposed as general
waste.
Parts List for 10.0 eV/Benzene PID Instruments
The table below lists spare parts for a 10.0 eV /benzene PID instrument.
Table 29: Spare Parts List for 10.0 eV/Benzene PID Instruments
Part Number
Description
13-0550
Lanyard for tube holder (for new style tube holder with knurled tip-breaker)
13-0700
Knurled tip-breaker on tube holder (for new style tube holder with knurled
tip-breaker)
13-0701
Black cap for knurled tip-breaker on tube holder (for new style tube holder
with knurled tip-breaker)
33-0563
PID sensor electrode stack, 10.0 eV/benzene sensor
33-7128
Filter tubes, 1 box, 10 tubes per box
33-7128-05
Filter tubes, 5 boxes, 10 tubes per box
51-1503
PID sensor replacement lamp, 10.0 eV/benzene sensor
80-0174
Tube holder
81-0100RK-03
Calibration cylinder, benzene, 5 ppm in air, 103 liter
81-0100RK-04
Calibration cylinder, benzene, 5 ppm in air, 34 liter aluminum
81-0104RK-03
Calibration cylinder, isobutylene, 10 ppm in air, 103 liter
81-0104RK-04
Calibration cylinder, isobutylene, 10 ppm in air, 34 liter aluminum
81-6XAX-DLV
Calibration kit: 34 liter 10 ppm IBL in air cylinder, demand flow regulator,
calibration tubing
81-6XZX
Calibration kit: 103 liter 5 ppm benzene in air cylinder, demand flow
regulator, calibration tubing
81-6XZX-DLV
Calibration kit: 34 liter 5 ppm benzene in air cylinder, demand flow
regulator, calibration tubing
82-0300RK
Aluminum oxide powder PID lamp cleaning kit, with electrode stack
removal tool, for 10.0 eV and 10.6 eV PID sensors’ lamps ONLY
PID-003L
PID sensor, 10.0 eV/benzene