GX-6000 Operator’s Manual
Appendix A: Maintenance Mode • 131
Appendix A: Maintenance Mode
Overview
This appendix describes the GX-6000 in Maintenance Mode. The GX-6000 is factory-set
to suit most applications. Update settings in Maintenance Mode only if required for your
specific application. Maintenance Mode items and their factory settings are listed in
Table 17 below.
Table 17: Maintenance Mode Menu Items
Menu Item (Page # of Description)
Choices or Action
Factory Setting
DATE (pg.137)
user-entered date/time
current date/time
DATE FORMAT (pg.137)
• DD/MM/YYYY
• MM/DD/YYYY
• YYYY/MM/DD
MM/DD/YYYY
GAS CALIBRATION (pg.138)
perform a calibration
N/A
BUMP TEST (pg.138)
perform a bump test
N/A
CAL SETTING
(pg.139)
CAL REMINDER (pg.139)
• ON
• OFF
ON
CAL INTERVAL (pg.139)
1-365 DAYS
90 DAYS
CAL EXPRD (pg.140)
• CONFIRM TO USE
• CANNOT USE
• NO EFFECT
CONFIRM TO
USE
CAL CHECK GAS (pg.140)
• 4 GAS
• ALL GAS
4 GAS
ESCAPE
return to main menu
N/A
BUMP
SETTING
(pg.141)
BUMP DISP (pg.141)
• ON
• OFF
OFF
BUMP
PARAMETERS
(pg.142)
GAS TIME
(pg.142)
• 30 SEC
• 45 SEC
• 60 SEC
• 90 SEC
30 SEC
CHECK
(pg.142)
• 10%
• 20%
• 30%
• 40%
• 50%
30%
CAL TIME
(pg.142)
• 90 SEC
• 120 SEC
90 SEC
AUTO CAL
(pg.143)
• ON
• OFF
ON