GX-6000 Operator’s Manual
Chapter 4: Calibration Mode • 79
22 . If you are using a 5-gas calibration cylinder:
a. Leave the tubing connected to the rigid tube on the probe. Continue with Step 23.
23 . Use the
▲
AIR and SHIFT
▼
(PANIC) buttons to move the cursor next to
GAS
SELECT
.
24 . Press and release the POWER/ENTER button. The first of the Gas Select Screens,
the standard four sensors screen, will be displayed. In addition to the standard four
sensors screen, there is a separate screen for each installed smart sensor, and an
ESCAPE
screen that will take you back to the Auto Cal menu. Use the
▲
AIR and
SHIFT
▼
(PANIC) buttons to scroll through the different screens.
AUTO
AUTO CAL
CONCENTRATION
GAS SELECT
ESCAPE
AUTO
GAS SELECT
CH4 50 %LEL
O2 12.0 %
H2S 25.0 ppm
CO 50 ppm
AUTO
GAS SELECT
VOC 100 ppm
AUTO
GAS SELECT
ESCAPE
▲
AIR or
SHIFT
▼
(PANIC)
▲
AIR or
SHIFT
▼
(PANIC)
▲
AIR or
SHIFT
▼
(PANIC)