158 • Appendix A: Maintenance Mode
GX-6000 Operator’s Manual
5 . Press and release POWER/ENTER. The Set Password Screen appears. 0000 is at the
bottom of the screen with the first 0 flashing.
6 . Use
▲
AIR or SHIFT
▼
(PANIC) to display a number from 0 to 9.
7 . Press and release POWER/ENTER to enter the selection and advance to the next
number.
8 . Repeat Step 6 and Step 7 to select the remaining numbers. When you press and
release POWER/ENTER to enter the last number, the password is saved and you
return to the main menu.
Restoring the Default Settings (RESTORE DEFAULT)
Each of the GX-6000 setup parameters, such as the auto calibration values, zero and span
settings, or parameters in Maintenance Mode, has a default setting. For the items in
Maintenance Mode, the default settings are the same as the standard factory settings. If
you want to return the GX-6000 to its default configuration, it is possible to do so by using
the Default Settings menu item in Maintenance Mode. Returning the GX-6000 to its
default configuration can be useful if various setup parameters have been changed in the
field and you want to return the GX-6000 to its original configuration as shipped from the
factory.
There are some special GX-6000 configurations that may have a different default
configuration than the standard. Consult RKI Instruments, Inc. for information regarding
non-standard default configurations.
WARNING:
When the GX-6000 is restored to its default configuration, the zero and
span values for each channel are reset. You must recalibrate all active
channels if you restore the GX-6000 to its default configuration.
1 . From the main menu, place the cursor in front of
RESTORE DEFAULT
.
MAINTE
MAINTE PASSWORD
INPUT
PASSWORD
0 0 0 0