104 • Chapter 6: Maintenance
GX-6000 Operator’s Manual
Replacing the Lithium Ion Battery Pack
NOTE:
Use Procell PC 1500 alkaline batteries or RKI Instruments, Inc. lithium-ion
battery pack 49-1619 to maintain the CSA classification of the GX-6000. Use
of other batteries will void the CSA classification and may void the warranty.
Do not mix old/new or different types of batteries.
NOTE:
Utiliser Procell 1500 piles alcalines ou RKI Instruments, Inc. pack batterie
lithium-ion 49-1619 de maintenir la classification CSA de la GX-6000.
L’utilisation d’autres piles annule la classification CSA et peut annuler la
garantie. Ne mélangez pas les anciennes/nouvelles ou différents types de piles.
1 . Turn off the GX-6000.
WARNING:
Do not remove the battery pack while the GX-6000 is on.
2 . Remove the rubber boot, if installed.
3 . Turn the GX-6000 over so that the flow chamber and battery cover are facing up and
the LCD is facing down.
4 . The battery pack release latch is located on the bottom of the instrument and is closer
to the front of the case than the back.
5 . Slide the battery pack release latch to the right with one hand and lift up on the
battery pack with the other.
Figure 14: Releasing the Battery Pack
6 . Slide the battery pack away from the instrument.
2.
1.
Lift Up on Battery Pack
Slide Battery Pack Release
Latch to the Right