Status Information, Error Messages and Troubleshooting
R&S
®
SMJ100A
819
Operating Manual 1403.7458.32 ─ 14
Error
Code
Error
Description
Remedy
242
Unknown list type
specified
The list type selected is not valid for
the required operation.
For instance, the file extension for
waveform list files is *.wv. It is not
possible to enter another file exten-
sion when selecting a list.
Check the selected list type.
261
Waveform Protected
The selected waveform file cannot be
transferred to a controller. The wave-
form is produced with simulation soft-
ware R&S WinIQSIM2 and is protec-
ted.
460
Cannot open file
The selected file can not be opened.
Check the path and file name.
461
Cannot write file
The file can not be written.
Check if the file is read-only.
462
Cannot read file
The file can not be read.
Check if the file contents are
compatible with the file type.
463
Filename missing
The required operation cannot be
executed because the file name is not
specified.
A file name has to be entered
when creating a new list.
464
Invalid filename exten-
sion
The file extension is not valid for the
required operation.
Check the file extension.
For instance, the file extension
for waveform list files is
*.wv
. It
is not possible to enter another
file extension when storing a list.
465
File contains invalid
data
The selected file contains data that is
not valid for the file type.
The file extension determines the
data that is valid for this file type. If
the file extension is changed the lists
are no longer recognized and the data
are therefore invalid.
Example: the extension of a wave-
form file (=
*.wv
) was changed to
*.txt
Check the file extension.
9.5 Resolving Network Connection Failures
Several issues may cause failures in the network connection to the instrument. This
section lists the most likely reasons and the recommended solutions.
Common reasons for network connection failures
●
Network connecting cables and cable connectors of poor quality
●
Incompatibility between the network interface of the R&S
SMJ and certain switches
or routers available on the market
●
An invalid IP address assigned to the instrument
Resolving Network Connection Failures
Содержание SMJ100A
Страница 34: ...Preface R S SMJ100A 18 Operating Manual 1403 7458 32 14 Notes on Screenshots...
Страница 86: ...Preparing for Use R S SMJ100A 70 Operating Manual 1403 7458 32 14 LXI Configuration...
Страница 139: ...Manual Operation R S SMJ100A 123 Operating Manual 1403 7458 32 14 Legend of Front Panel Controls...
Страница 140: ...Manual Operation R S SMJ100A 124 Operating Manual 1403 7458 32 14 Legend of Front Panel Controls...
Страница 454: ...Instrument Function R S SMJ100A 438 Operating Manual 1403 7458 32 14 Baseband Signal Baseband Block...
Страница 826: ...Remote Control Commands R S SMJ100A 810 Operating Manual 1403 7458 32 14 UNIT Subsystem...
Страница 828: ...Maintenance R S SMJ100A 812 Operating Manual 1403 7458 32 14 Storing and Packing...
Страница 843: ...Hardware Interfaces R S SMJ100A 827 Operating Manual 1403 7458 32 14 Figure A 2 BERT circuit diagram BERT Connector...
Страница 844: ...Hardware Interfaces R S SMJ100A 828 Operating Manual 1403 7458 32 14 BERT Connector...