Remote Control Commands
R&S
®
SMJ100A
644
Operating Manual 1403.7458.32 ─ 14
[:SOURce<hw>]:BB:DM:TRIGger:OUTPut<ch>:PULSe:FREQuency?
Queries the pulse frequency of the pulsed marker signal in the setting
SOURce:BB:DM:TRIGger:OUTPut:MODE PULSe
. The pulse frequency is derived by
dividing the symbol rate by the divider.
Return values:
<Frequency>
float
Example:
BB:DM:TRIG:OUTP2:PULS:DIV 2
sets the divider marker signal on output MARKER 2 to the value
2.
BB:DM:TRIG:OUTP2:MODE PULS
enables the pulsed marker signal.
BB:DM:TRIG:OUTP2:PULS:FREQ?
queries the pulse frequency of the marker signal.
Response: 33 000
the resulting pulse frequency is
Usage:
Query only
Filter Settings
[:SOURce<hw>]:BB:DM:FILTer:OSAMpling
...................................................................... 644
[:SOURce<hw>]:BB:DM:FILTer:OSAMpling:AUTO
............................................................ 645
[:SOURce<hw>]:BB:DM:FILTer:PARameter:COSine[:ROLLoff]
...........................................645
[:SOURce<hw>]:BB:DM:FILTer:PARameter:GAUSs
..........................................................645
[:SOURce<hw>]:BB:DM:FILTer:PARameter:LPASs
...........................................................645
[:SOURce<hw>]:BB:DM:FILTer:PARameter:LPASSEVM
................................................... 645
[:SOURce<hw>]:BB:DM:FILTer:PARameter:PGAuss
.........................................................645
[:SOURce<hw>]:BB:DM:FILTer:PARameter:RCOSine
....................................................... 645
[:SOURce<hw>]:BB:DM:FILTer:PARameter:SPHase
.........................................................645
[:SOURce<hw>]:BB:DM:FILTer:PARameter:APCO25
........................................................645
[:SOURce<hw>]:BB:DM:FILTer:PARameter:COSine:BANDwidth
........................................646
[:SOURce<hw>]:BB:DM:FILTer:PARameter:APCO25Lsm:GAUSs
...................................... 646
[:SOURce<hw>]:BB:DM:FILTer:PARameter:APCO25Lsm:LOWPass
.................................. 646
[:SOURce<hw>]:BB:DM:FILTer:TYPE
..............................................................................647
[:SOURce<hw>]:DM:FILTer:PARameter
...........................................................................647
[:SOURce<hw>]:BB:DM:FILTer:OSAMpling
<OSampling>
The command sets the upsampling factor.
Parameters:
<OSampling>
integer
Range:
1 to 32
*RST:
32
Example:
BB:DM:FILT:OSAM 32
sets the upsampling factor to 32.
Manual operation:
See
SOURce Subsystem
Содержание SMJ100A
Страница 34: ...Preface R S SMJ100A 18 Operating Manual 1403 7458 32 14 Notes on Screenshots...
Страница 86: ...Preparing for Use R S SMJ100A 70 Operating Manual 1403 7458 32 14 LXI Configuration...
Страница 139: ...Manual Operation R S SMJ100A 123 Operating Manual 1403 7458 32 14 Legend of Front Panel Controls...
Страница 140: ...Manual Operation R S SMJ100A 124 Operating Manual 1403 7458 32 14 Legend of Front Panel Controls...
Страница 454: ...Instrument Function R S SMJ100A 438 Operating Manual 1403 7458 32 14 Baseband Signal Baseband Block...
Страница 826: ...Remote Control Commands R S SMJ100A 810 Operating Manual 1403 7458 32 14 UNIT Subsystem...
Страница 828: ...Maintenance R S SMJ100A 812 Operating Manual 1403 7458 32 14 Storing and Packing...
Страница 843: ...Hardware Interfaces R S SMJ100A 827 Operating Manual 1403 7458 32 14 Figure A 2 BERT circuit diagram BERT Connector...
Страница 844: ...Hardware Interfaces R S SMJ100A 828 Operating Manual 1403 7458 32 14 BERT Connector...