Instrument Function
R&S
®
SMJ100A
147
Operating Manual 1403.7458.32 ─ 14
Remote command:
:SYSTem:COMMunicate:GPIB[:SELF]:ADDRess
Visa Resource Strings
Displays the visa resource strings, used for remote control of the instrument. Each
interface requires an individual unique address, to identify the instrument for remote
control.
Remote command:
:SYSTem:COMMunicate:HISLip:RESource?
:SYSTem:COMMunicate:NETWork:RESource?
:SYSTem:COMMunicate:SOCKet:RESource?
:SYSTem:COMMunicate:GPIB:RESource?
Goto Local
Switches the instrument to operate in local control mode.
Switching from remote to local control mode can be also done with one of the following
actions:
●
manually with the LOCAL key on the front panel
●
with the interface command
>L
via the remote control interface
●
with the key combination CTRL + Q.
Remote command:
>L
5.2.3.14
LXI Status
The "LXI - LAN eXtensions for Instruments - Status..." dialog displays the settings and
status of the LAN and allows to reset the LAN connection.
For more information on LXI, see
Chapter 2.9, "LXI Configuration"
LAN Status
The LED indicates the LXI status.
The LAN Status is also indicated with the LED "LAN Status" on the front panel of the
instrument.
"green"
normal operation
General Instrument Settings
Содержание SMJ100A
Страница 34: ...Preface R S SMJ100A 18 Operating Manual 1403 7458 32 14 Notes on Screenshots...
Страница 86: ...Preparing for Use R S SMJ100A 70 Operating Manual 1403 7458 32 14 LXI Configuration...
Страница 139: ...Manual Operation R S SMJ100A 123 Operating Manual 1403 7458 32 14 Legend of Front Panel Controls...
Страница 140: ...Manual Operation R S SMJ100A 124 Operating Manual 1403 7458 32 14 Legend of Front Panel Controls...
Страница 454: ...Instrument Function R S SMJ100A 438 Operating Manual 1403 7458 32 14 Baseband Signal Baseband Block...
Страница 826: ...Remote Control Commands R S SMJ100A 810 Operating Manual 1403 7458 32 14 UNIT Subsystem...
Страница 828: ...Maintenance R S SMJ100A 812 Operating Manual 1403 7458 32 14 Storing and Packing...
Страница 843: ...Hardware Interfaces R S SMJ100A 827 Operating Manual 1403 7458 32 14 Figure A 2 BERT circuit diagram BERT Connector...
Страница 844: ...Hardware Interfaces R S SMJ100A 828 Operating Manual 1403 7458 32 14 BERT Connector...