Remote Control Commands
R&S
®
SMJ100A
506
Operating Manual 1403.7458.32 ─ 14
Figure 7-7: Contents of SICO.txt: first column Sine (I), second column Cosine (Q)
2. Convert the values from the file
SICO.txt
into binary format consisting of 16-bit
signed integer numbers. The numeric range between –1.0 and +1.0 corresponds to
the modulation range of the waveform 16-bit D/A converter of -32767 to +32767.
+1.0 ->
32767 ->
= 0x7FFF
0.0 ->
0 ->
= 0x0000
-1.0 ->
-32767 -> = 0x8001
The
shows the calculation and conversion steps. The highlighted col-
umns contain the resulting I and Q values represented in Little endian format.