Instrument Function
R&S
®
SMJ100A
238
Operating Manual 1403.7458.32 ─ 14
Filter
Determines the length of the filter used for the measurement. The filter length affects
the measurement time directly.
The averaging filter is used to reduce fluctuations in the measured result to the extent
desired. Such fluctuations can be caused by inherent noise of the measuring instru-
ment, modulation of the measurement signal or beats from the superposition of adja-
cent carriers. A more stable display has to be traded off against longer measurements.
The measurement result is obtained from a two-stage averaging process.
Note:
Longer measurements do not mean that it takes longer to display a new result,
but rather that it takes longer for the result to settle when the power changes.
Measurements are continuously repeated in a predefined time window. The measure-
ment result is obtained by averaging the measured values for the last 2N time win-
dows. The number N is the filter length, the factor of 2 arises because the output sig-
nals from the microwave detector to suppress low-frequency noise are chopped at the
same rate as the time windows, which means that an independent measured value can
only be obtained from two consecutive values. As the filter length is the multiplier for
the time window it directly influences the measurement time.
The filter length can be selected automatically or can be manually set to a fixed value.
As a preliminary, you should always check if the auto mode is giving satisfactory
results because you will always have to adjust an optimal, manual filter-length setting if
the power is not constant.
Selection "Fixed Noise" is offered for achieving defined measurement accuracy.
"Auto"
The filter length is automatically selected and adapted to the currently
measured value. With very high signals the filter length and therefore
the measurement time can be short. With very low signal levels the
filter length and therefore the measurement time is increased in order
to reduce noise. The used filter length is indicated in the field to the
right, see
.
"User"
The filter length is set manually.
The filter length is entered in the entry window to the right. As the fil-
ter length works as a multiplier for the time window, this results in a
constant measurement time.
Note:
The time window varies depending on the used sensor. For
most sensors it is fixed to 20 ms. For the R&S NRP-Z81 sensor it is
10 us. Therefore, the user filter length for the R&S NRP-Z81 has be
about 1000 times larger than the filter length for other sensors in
order to achieve the same filtering result.
The
button can be used to search for the optimum filter
length for the current measurement conditions. The found filter length
is indicated in the field to the right, see
"Fixed Noise"
The averaging factor is selected so that the sensors intrinsic noise (2
standard deviations) does not exceed the specified noise content.
The desired noise content is entered in the entry field to the right, see
To avoid very long settling times when the power is low, the averag-
ing factor can be limited with the
RF Signal and Analog Modulations - RF Block
Содержание SMJ100A
Страница 34: ...Preface R S SMJ100A 18 Operating Manual 1403 7458 32 14 Notes on Screenshots...
Страница 86: ...Preparing for Use R S SMJ100A 70 Operating Manual 1403 7458 32 14 LXI Configuration...
Страница 139: ...Manual Operation R S SMJ100A 123 Operating Manual 1403 7458 32 14 Legend of Front Panel Controls...
Страница 140: ...Manual Operation R S SMJ100A 124 Operating Manual 1403 7458 32 14 Legend of Front Panel Controls...
Страница 454: ...Instrument Function R S SMJ100A 438 Operating Manual 1403 7458 32 14 Baseband Signal Baseband Block...
Страница 826: ...Remote Control Commands R S SMJ100A 810 Operating Manual 1403 7458 32 14 UNIT Subsystem...
Страница 828: ...Maintenance R S SMJ100A 812 Operating Manual 1403 7458 32 14 Storing and Packing...
Страница 843: ...Hardware Interfaces R S SMJ100A 827 Operating Manual 1403 7458 32 14 Figure A 2 BERT circuit diagram BERT Connector...
Страница 844: ...Hardware Interfaces R S SMJ100A 828 Operating Manual 1403 7458 32 14 BERT Connector...