Instrument Function
R&S
®
SMJ100A
243
Operating Manual 1403.7458.32 ─ 14
State
Activates amplitude modulation.
Activation of AM deactivates:
●
BB AM
●
I/Q modulation
●
arbitrary waveform generation
●
digital modulation
●
digital standards
Remote command:
AM Source
Selects the source for the AM modulation signal.
"Internal"
Uses the internal LF generator as modulation signal source for AM.
"External"
Uses an externally applied modulation signal.
The external signal is input via the EXT MOD connector.
" Extern"
Uses both, the internal and externally applied modulation signal, for
example to perform two-tone AM.
Remote command:
AM Depth
Sets the modulation depth in percent.
Remote command:
LF Gen Frequency
Sets the frequency of the LF generator.
This setting affects all analog modulations which use the LF generator as the internal
modulation source.
Remote command:
RF Signal and Analog Modulations - RF Block
Содержание SMJ100A
Страница 34: ...Preface R S SMJ100A 18 Operating Manual 1403 7458 32 14 Notes on Screenshots...
Страница 86: ...Preparing for Use R S SMJ100A 70 Operating Manual 1403 7458 32 14 LXI Configuration...
Страница 139: ...Manual Operation R S SMJ100A 123 Operating Manual 1403 7458 32 14 Legend of Front Panel Controls...
Страница 140: ...Manual Operation R S SMJ100A 124 Operating Manual 1403 7458 32 14 Legend of Front Panel Controls...
Страница 454: ...Instrument Function R S SMJ100A 438 Operating Manual 1403 7458 32 14 Baseband Signal Baseband Block...
Страница 826: ...Remote Control Commands R S SMJ100A 810 Operating Manual 1403 7458 32 14 UNIT Subsystem...
Страница 828: ...Maintenance R S SMJ100A 812 Operating Manual 1403 7458 32 14 Storing and Packing...
Страница 843: ...Hardware Interfaces R S SMJ100A 827 Operating Manual 1403 7458 32 14 Figure A 2 BERT circuit diagram BERT Connector...
Страница 844: ...Hardware Interfaces R S SMJ100A 828 Operating Manual 1403 7458 32 14 BERT Connector...