Remote Control Commands
R&S
®
SMJ100A
788
Operating Manual 1403.7458.32 ─ 14
Example:
SWE:RES
resets all active sweeps to the starting point.
Usage:
Event
Manual operation:
See
7.17 STATus Subsystem
This system contains the commands for the status reporting system. See also
ter 6.5, "Status Reporting System"
on page 473 for detailed information.
on page 486 has no effect on the status registers.
Value ranges
●
Queries return the current value of the respective register, which permits a check of
the device status.
Return values: A decimal value in the range 0 to 32767 (=2
15
-1)
●
The configuration commands set the respective register thus determining which
status changes of the R&S
SMJ cause the status registers to be changed.
Setting values: A decimal value in the range 0 to 32767 (=2
15
-1)
...................................................................................... 788
.......................................................................................... 789
.......................................................................................... 789
....................................................................................789
.................................................................................... 789
:STATus:QUEStionable:CONDition
..................................................................................790
...................................................................................... 790
......................................................................................790
:STATus:QUEStionable:NTRansition
................................................................................791
:STATus:QUEStionable:PTRansition
................................................................................791
............................................................................................... 791
:STATus:OPERation:CONDition
<Condition>
Sets the content of the CONDition part of the STATus:OPERation register. This part
contains information on the action currently being performed in the instrument. The
content is not deleted after being read out because it indicates the current hardware
status.
Parameters:
<Condition>
string
Example:
:STATus:OPERation:CONDition?
queries the Status:Operation:Condition register.
STATus Subsystem
Содержание SMJ100A
Страница 34: ...Preface R S SMJ100A 18 Operating Manual 1403 7458 32 14 Notes on Screenshots...
Страница 86: ...Preparing for Use R S SMJ100A 70 Operating Manual 1403 7458 32 14 LXI Configuration...
Страница 139: ...Manual Operation R S SMJ100A 123 Operating Manual 1403 7458 32 14 Legend of Front Panel Controls...
Страница 140: ...Manual Operation R S SMJ100A 124 Operating Manual 1403 7458 32 14 Legend of Front Panel Controls...
Страница 454: ...Instrument Function R S SMJ100A 438 Operating Manual 1403 7458 32 14 Baseband Signal Baseband Block...
Страница 826: ...Remote Control Commands R S SMJ100A 810 Operating Manual 1403 7458 32 14 UNIT Subsystem...
Страница 828: ...Maintenance R S SMJ100A 812 Operating Manual 1403 7458 32 14 Storing and Packing...
Страница 843: ...Hardware Interfaces R S SMJ100A 827 Operating Manual 1403 7458 32 14 Figure A 2 BERT circuit diagram BERT Connector...
Страница 844: ...Hardware Interfaces R S SMJ100A 828 Operating Manual 1403 7458 32 14 BERT Connector...