Instrument Function
R&S
®
SMJ100A
232
Operating Manual 1403.7458.32 ─ 14
To increase the level accuracy and provide precise measurement results, the instru-
ment considers the changes in VSWR for the different attenuator settings. It automati-
cally selects the corresponding reference level for each of the attenuator steps. For
ranges from 40 dB on, the changes in VSWR are negligible; it is thus sufficient to apply
the same correction value as for 35 dB attenuation.
The acquired measurement data is stored in a multi level correction table, containing
one row per each frequency and ten additional power columns.
1
= single level measurement power correction at the output level
2
= "Power [1]" power correction at high power attenuator settings
3a .. 3h = "Power [2] .. Power [9]" power corrections at the different attenuator steps
The R&S
SMJ performs the multi level correction measurements in addition to a single
level correction measurement. This enables you, with one single user correction list, to
switch between the single and the multi level correction results.
Multi Level Values
The measurement values for the different attenuator settings are retrieved separately.
The numeric suffix at MLEV correlates to the index of the column, e.g. suffix 3 denotes
the column with the values for attenuator setting 5 dB.
Remote command:
[:SOURce<hw>]:CORRection:CSET:DATA:POWer:MLEVel:STATe
[:SOURce<hw>]:CORRection:CSET:DATA:POWer:MLEVel<ch>?
5.5.9 Reverse Power Protection
Starting from serial number 101001, the R&S
SMJ (all frequency options) is equipped
against overloading by an external signal applied to the RF output.
The reverse power protection is tripped when the power of the external signal becomes
too high. A relay opens and interrupts the internal connection to the RF output. This
condition is indicated in the display header by the "OVERLOAD" status message.
Overload
If an "Overload" status message is indicated in the display header, reset the overload
protection by pressing the RF ON/OFF key.
The RF input is activated when the overload protection is reset.
Remote command:
:OUTPut<hw>:PROTection:TRIPped?
RF Signal and Analog Modulations - RF Block
Содержание SMJ100A
Страница 34: ...Preface R S SMJ100A 18 Operating Manual 1403 7458 32 14 Notes on Screenshots...
Страница 86: ...Preparing for Use R S SMJ100A 70 Operating Manual 1403 7458 32 14 LXI Configuration...
Страница 139: ...Manual Operation R S SMJ100A 123 Operating Manual 1403 7458 32 14 Legend of Front Panel Controls...
Страница 140: ...Manual Operation R S SMJ100A 124 Operating Manual 1403 7458 32 14 Legend of Front Panel Controls...
Страница 454: ...Instrument Function R S SMJ100A 438 Operating Manual 1403 7458 32 14 Baseband Signal Baseband Block...
Страница 826: ...Remote Control Commands R S SMJ100A 810 Operating Manual 1403 7458 32 14 UNIT Subsystem...
Страница 828: ...Maintenance R S SMJ100A 812 Operating Manual 1403 7458 32 14 Storing and Packing...
Страница 843: ...Hardware Interfaces R S SMJ100A 827 Operating Manual 1403 7458 32 14 Figure A 2 BERT circuit diagram BERT Connector...
Страница 844: ...Hardware Interfaces R S SMJ100A 828 Operating Manual 1403 7458 32 14 BERT Connector...