Remote Control Commands
R&S
®
SMJ100A
777
Operating Manual 1403.7458.32 ─ 14
Parameters:
<Source>
INTernal | EXTernal | ELOop
INTernal
The internal reference oscillator is used.
EXTernal
An external reference signal is used. It must be input at the REF
IN connector at the rear of the instrument.
The instrument is informed of the frequency of the external refer-
ence signal by means of the command
ROSCillator:EXTernal:FREQuency
*RST:
---
Example:
ROSC:SOUR EXT
selects the external source.
ROSC:EXT:FREQ 5 MHz
informs the instrument that the external reference has a fre-
quency of 5 MHz.
Manual operation:
See
7.16.18 SOURce:SWEep Subsystem
The
SOURce:
subsystem contains the commands for configuring RF sweep signals.
●
The keyword
[:FREQuency]
can be ommitted, then the commands are SCPI-
compliant.
●
To activate a RF sweep mode, use the following commands:
–
RF frequency sweep:
SOURce:FREQuency:MODE SWEep
(
SOURce:FREQuency:MODE CW
(off))
–
RF level sweep:
SOURce:POWer:MODE SWEep
(
SOURce:POWer:MODE CW
(off))
●
All sweeps, including the LF sweep, can be set independently from each other.
This example shows how to set up a frequency sweep.
1. Set the sweep range.
[SOURce:]FREQuency:CENTer 200 MHz
[SOURce:]FREQuency:SPAN 300 MHz
2. Select linear or logarithmic spacing.
[SOURce:]SWEep[:FREQuency]:SPACing LIN
3. Set the step width and dwell time.
[SOURce:]SWEep[:FREQuency]:STEP:LINear 20 MHz
[SOURce:]SWEep[:FREQuency]:DWELl 12 ms
4. Select the trigger mode.
TRIGger:]FSWeep:SOURce SINGle
SOURce Subsystem
Содержание SMJ100A
Страница 34: ...Preface R S SMJ100A 18 Operating Manual 1403 7458 32 14 Notes on Screenshots...
Страница 86: ...Preparing for Use R S SMJ100A 70 Operating Manual 1403 7458 32 14 LXI Configuration...
Страница 139: ...Manual Operation R S SMJ100A 123 Operating Manual 1403 7458 32 14 Legend of Front Panel Controls...
Страница 140: ...Manual Operation R S SMJ100A 124 Operating Manual 1403 7458 32 14 Legend of Front Panel Controls...
Страница 454: ...Instrument Function R S SMJ100A 438 Operating Manual 1403 7458 32 14 Baseband Signal Baseband Block...
Страница 826: ...Remote Control Commands R S SMJ100A 810 Operating Manual 1403 7458 32 14 UNIT Subsystem...
Страница 828: ...Maintenance R S SMJ100A 812 Operating Manual 1403 7458 32 14 Storing and Packing...
Страница 843: ...Hardware Interfaces R S SMJ100A 827 Operating Manual 1403 7458 32 14 Figure A 2 BERT circuit diagram BERT Connector...
Страница 844: ...Hardware Interfaces R S SMJ100A 828 Operating Manual 1403 7458 32 14 BERT Connector...