Remote Control Commands
R&S
®
SMJ100A
582
Operating Manual 1403.7458.32 ─ 14
In the case of an external clock source (selection
ARB:CLOCk:SOURce EXTernal
)
the clock for the external source must be entered with this command.
Parameters:
<Clock>
float
Increment: 0.001
Example:
BB:ARB:CLOC:SOUR INT
selects the internal clock source for generating waveforms.
BB:ARB:CLOC 0.5 MHz
sets the clock rate to 0.5 MHz.
Manual operation:
See
[:SOURce<hw>]:BB:ARBitrary:CLOCk:MODE
<Mode>
The command enters the type of externally supplied clock (
:BB:ARB:CLOCk:SOURce
EXTernal
). When MSAMple is used, a multiple of the sample clock is supplied via the
CLOCK connector and the sample clock is derived internally from this. The multiplier is
entered with the command
:BB:ARB:CLOCk:MULTiplier
.
Parameters:
<Mode>
SAMPle | MSAMple
*RST:
SAMPle
Example:
BB:ARB:CLOC:MODE SAMP
selects clock type "Sample", i.e. the supplied clock is a sample
clock.
Manual operation:
See
[:SOURce<hw>]:BB:ARBitrary:CLOCk:MULTiplier
<Multiplier>
The command specifies the multiplier for clock type "Multiple Samples"
(
:BB:ARB:CLOCk:MODE MSAM
) in the case of an external clock source.
Parameters:
<Multiplier>
integer
Range:
1 to 64
*RST:
4
Example:
BB:ARB:CLOC:SOUR EXT
selects the external clock source. The clock is supplied via the
CLOCK connector.
BB:ARB:CLOC:MODE MSAM
selects clock type Multiple Samples, i.e. the supplied clock has a
rate which is a multiple of the sample rate.
BB:ARB:CLOC:MULT 12
The multiplier for the external clock rate is 12.
Manual operation:
See
SOURce Subsystem
Содержание SMJ100A
Страница 34: ...Preface R S SMJ100A 18 Operating Manual 1403 7458 32 14 Notes on Screenshots...
Страница 86: ...Preparing for Use R S SMJ100A 70 Operating Manual 1403 7458 32 14 LXI Configuration...
Страница 139: ...Manual Operation R S SMJ100A 123 Operating Manual 1403 7458 32 14 Legend of Front Panel Controls...
Страница 140: ...Manual Operation R S SMJ100A 124 Operating Manual 1403 7458 32 14 Legend of Front Panel Controls...
Страница 454: ...Instrument Function R S SMJ100A 438 Operating Manual 1403 7458 32 14 Baseband Signal Baseband Block...
Страница 826: ...Remote Control Commands R S SMJ100A 810 Operating Manual 1403 7458 32 14 UNIT Subsystem...
Страница 828: ...Maintenance R S SMJ100A 812 Operating Manual 1403 7458 32 14 Storing and Packing...
Страница 843: ...Hardware Interfaces R S SMJ100A 827 Operating Manual 1403 7458 32 14 Figure A 2 BERT circuit diagram BERT Connector...
Страница 844: ...Hardware Interfaces R S SMJ100A 828 Operating Manual 1403 7458 32 14 BERT Connector...