![R&S SMJ100A Скачать руководство пользователя страница 328](http://html1.mh-extra.com/html/rands/smj100a/smj100a_operating-manual_3086611328.webp)
Instrument Function
R&S
®
SMJ100A
312
Operating Manual 1403.7458.32 ─ 14
When activated, information on the peak power, power and crest factor are indicated.
Additionally the duration of the measurement can be set.
Remote command:
[:SOURce<hw>]:BB:MEASurement[:STATe]
Signal Source - Digital Impairments - Measurement
Selects the signal for the measurement. The power and crest factor can be retrieved
either from the interference-free digital I/Q baseband signal or the signal superimposed
by the noise generator AWGN.
"Baseband w/o
AWGN"
Activates the measurement of the interference-free baseband signal.
"Baseband
with AWGN"
Activates the measurement of the baseband signal after an AWGN
noise signal is added.
Remote command:
[:SOURce<hw>]:BB:MEASurement:SOURce
Duration - Digital Impairments - Measurement
Enters the time for the measurement of the digital I/Q baseband signal in the range of
1 s ... 3 h. The time is to be entered in format "hh:mm:ss".
Remote command:
[:SOURce<hw>]:BB:MEASurement:DURation
Remaining Time - Digital Impairments - Measurement
Indicates the remaining time until the measurement of the digital I/Q baseband signal is
completed. The time is displayed in format "hh:mm:ss".
Remote command:
[:SOURce<hw>]:BB:MEASurement:RTIMe?
Crest Factor - Digital Impairments - Measurement
Indicates the crest factor of the baseband signal during measurement.
Remote command:
[:SOURce<hw>]:BB:MEASurement:CFACtor?
Peak Power - Digital Impairments - Measurement
Indicates the peak power of the baseband signal relative to full scale of 0.5 V (in terms
of dB full scale) during measurement.
Remote command:
[:SOURce<hw>]:BB:MEASurement:POWer:PEAK?
Power - Digital Impairments - Measurement
Indicates the power of the baseband signal relative to full scale of 0.5 V (in terms of dB
full scale) during measurement.
Remote command:
[:SOURce<hw>]:BB:MEASurement:POWer:RMS?
Impairment of Digital I/Q Signal and Noise Generator - AWGN/IMP Block
Содержание SMJ100A
Страница 34: ...Preface R S SMJ100A 18 Operating Manual 1403 7458 32 14 Notes on Screenshots...
Страница 86: ...Preparing for Use R S SMJ100A 70 Operating Manual 1403 7458 32 14 LXI Configuration...
Страница 139: ...Manual Operation R S SMJ100A 123 Operating Manual 1403 7458 32 14 Legend of Front Panel Controls...
Страница 140: ...Manual Operation R S SMJ100A 124 Operating Manual 1403 7458 32 14 Legend of Front Panel Controls...
Страница 454: ...Instrument Function R S SMJ100A 438 Operating Manual 1403 7458 32 14 Baseband Signal Baseband Block...
Страница 826: ...Remote Control Commands R S SMJ100A 810 Operating Manual 1403 7458 32 14 UNIT Subsystem...
Страница 828: ...Maintenance R S SMJ100A 812 Operating Manual 1403 7458 32 14 Storing and Packing...
Страница 843: ...Hardware Interfaces R S SMJ100A 827 Operating Manual 1403 7458 32 14 Figure A 2 BERT circuit diagram BERT Connector...
Страница 844: ...Hardware Interfaces R S SMJ100A 828 Operating Manual 1403 7458 32 14 BERT Connector...