Chapter 5
5-26
V60 Ventilator Service Manual
1049766 Rev A
Diagnostic Mode and Troubleshooting
120E
Alarm message:
Low Minute Ventilation
Estimated minute ventilation is less than the
Lo
.
V
E
setting. Escalates to a high-priority alarm if the alarm
condition persists for more than 60 sec.
Check patient. Confirm that
ventilator and alarm settings are
appropriate. If problem persists,
provide alternative ventilation.
Service ventilator.
120F
Alarm message:
Low Tidal Volume
Estimated tidal volume is less than the Lo V
T
setting.
Escalates to a high-priority alarm if the alarm condition
persists for more than 60 sec.
Check patient. Confirm that
ventilator and alarm settings are
appropriate. If problem persists,
provide alternative ventilation.
Service ventilator.
1210 Alarm
message:
High Tidal Volume
Measured estimated tidal volume is greater than the Hi V
T
setting. Escalates to a high-priority alarm if the alarm
condition persists for more than 60 sec.
Check patient. Confirm that
ventilator and alarm settings are
appropriate. If problem persists,
provide alternative ventilation.
Service ventilator.
1211 Alarm
message:
High Rate
Measured respiratory rate is greater than the Hi Rate
setting. Escalates to a high-priority alarm if the alarm
condition persists for more than 60 sec.
Check patient. Confirm that
ventilator and alarm settings are
appropriate. If problem persists,
provide alternative ventilation.
Service ventilator.
1212
Alarm message:
Running on Internal Battery
System is powered by internal battery. Autoresets when
ventilator is connected to AC power.
Connect ventilator to AC power.
1213
Alarm message:
Using Default Settings
Displayed after power on if setting values are corrupted or
not set, or if default values were restored by the user.
Check patient. Check and adjust
settings as required.
1214 Alarm
message:
AVAPS: Target Vt Not Achieved.
Insufficient Max Pressure
AVAPS target pressure exceeds AVAPS Max P setting. The
ventilator limits applied pressure to Max P.
Check patient. Confirm that
pressure settings are compatible
with target. Evaluate pressure and
volume settings.
1215
Alarm message:
AVAPS: Target Vt Exceeded. Min Pressure
Too High
AVAPS target pressure is less than Min P setting. The
ventilator limits applied pressure to Min P.
Check patient. Confirm that
pressure settings are compatible
with target. Evaluate pressure and
volume settings.
Table 5-1: Diagnostic Codes and Troubleshooting
Diagnostic
code
Description
Recommended repair
Содержание Respironics V60
Страница 1: ...Respironics V60 Ventilator Service Manual ...
Страница 8: ...viii V60 Ventilator Service Manual 1049766 Rev A Table of Contents This page is intentionally blank ...
Страница 204: ...Chapter 8 8 50 V60 Ventilator Service Manual 1049766 Rev A Performance Verification This page is intentionally blank ...
Страница 220: ...Chapter 9 9 16 V60 Ventilator Service Manual 1049766 Rev A Replacement Parts List This page is intentionally blank ...
Страница 242: ...Appendix A A 14 V60 Ventilator Service Manual 1049766 Rev A This page is intentionally blank ...
Страница 244: ...Appendix B B 2 V60 Ventilator Service Manual 1049766 Rev A This page is intentionally blank ...