Chapter 8
Performance Verification
1049766 Rev A
V60 Ventilator Service Manual
8-23
Figure 8-16: Diagnostics Menu - Pneumatics Tab Selected
7. Set ventilator pressure values and verify that they correspond to
pressure measurements:
NOTE:
Add the machine pressure offset reading (pneumatics screen) at 0 to the
analyzer readings for 1, 35, and 60 cmH
2
O.
8. To reset the pressure setting to 0, touch
Flow
then
Accept
.
9. The pressure accuracy test is complete.
Remove low-pressure plug if
additional testing is not required
.
Set
Pressure
to:
Verify that analyzer, average
Machine
, and
Proximal
displays read:
0 cmH
2
O
-1.5 to 1.5 cmH
2
O
1 cmH
2
O
0.5 to 1.5 cmH
2
O
35 cmH
2
O
32.5 to 38.5 cmH
2
O
60 cmH
2
O
54 to 66 cmH
2
O
Содержание Respironics V60
Страница 1: ...Respironics V60 Ventilator Service Manual ...
Страница 8: ...viii V60 Ventilator Service Manual 1049766 Rev A Table of Contents This page is intentionally blank ...
Страница 204: ...Chapter 8 8 50 V60 Ventilator Service Manual 1049766 Rev A Performance Verification This page is intentionally blank ...
Страница 220: ...Chapter 9 9 16 V60 Ventilator Service Manual 1049766 Rev A Replacement Parts List This page is intentionally blank ...
Страница 242: ...Appendix A A 14 V60 Ventilator Service Manual 1049766 Rev A This page is intentionally blank ...
Страница 244: ...Appendix B B 2 V60 Ventilator Service Manual 1049766 Rev A This page is intentionally blank ...