Chapter 7
Component Removal/Installation
1049766 Rev A
V60 Ventilator Service Manual
7-15
7.14 Speakers
Follow these steps to remove the alarm speakers (Figure 7-14). Reverse to
install.
1. Remove front panel (section 7.10).
2. Disconnect speaker #2 from the power management (PM) PCBA.
(Speaker #1 is disconnected when the MC PCBA is disconnected.)
3. Remove the screw and washer that attaches each speaker to the base.
Figure 7-14: Removing the Alarm Speakers
Speaker #2
(to PM PCBA)
Speaker #1
(to MC PCBA)
M3 x 6 screw (x2)
(62 + 5.7 N cm/
5.5 + 0.5 in.-lbf)
M3 washer (x2)
Содержание Respironics V60
Страница 1: ...Respironics V60 Ventilator Service Manual ...
Страница 8: ...viii V60 Ventilator Service Manual 1049766 Rev A Table of Contents This page is intentionally blank ...
Страница 204: ...Chapter 8 8 50 V60 Ventilator Service Manual 1049766 Rev A Performance Verification This page is intentionally blank ...
Страница 220: ...Chapter 9 9 16 V60 Ventilator Service Manual 1049766 Rev A Replacement Parts List This page is intentionally blank ...
Страница 242: ...Appendix A A 14 V60 Ventilator Service Manual 1049766 Rev A This page is intentionally blank ...
Страница 244: ...Appendix B B 2 V60 Ventilator Service Manual 1049766 Rev A This page is intentionally blank ...