Chapter 8
Performance Verification
1049766 Rev A
V60 Ventilator Service Manual
8-9
8.4 Pneumatic
Calibration Analyzer
Setup
Follow these steps to set up the Certifier FA Plus pneumatic calibration
analyzer, including:
•
Measurement Selection Screen
•
Averaging Setup Menu
•
Trigger Options menu
•
Configurations menu
NOTE:
When using a pneumatic calibration analyzer, remember the analyzer
should be in STP when measuring flow.
8.4.1
Measurement Selection Screen
Use the Measurement Selection Screen to add or remove parameters from the
Certifier FA Plus screen.
1. Touch any parameter on the analyzer touch screen to display the
Measurement Selection Screen (Figure 8-2).
Figure 8-2: Measurement Selection Screen (Certifier FA Plus)
2. To add a parameter, touch the parameter name in the left window then
touch the right arrow.
To remove a parameter, touch the parameter name in the right window
then touch the left arrow.
Use the up and down arrows to determine the order of the selected
parameters.
Recommended parameters:
•
High Pressure (psig)
Содержание Respironics V60
Страница 1: ...Respironics V60 Ventilator Service Manual ...
Страница 8: ...viii V60 Ventilator Service Manual 1049766 Rev A Table of Contents This page is intentionally blank ...
Страница 204: ...Chapter 8 8 50 V60 Ventilator Service Manual 1049766 Rev A Performance Verification This page is intentionally blank ...
Страница 220: ...Chapter 9 9 16 V60 Ventilator Service Manual 1049766 Rev A Replacement Parts List This page is intentionally blank ...
Страница 242: ...Appendix A A 14 V60 Ventilator Service Manual 1049766 Rev A This page is intentionally blank ...
Страница 244: ...Appendix B B 2 V60 Ventilator Service Manual 1049766 Rev A This page is intentionally blank ...